Verifiering av EMC-kammare – Analys av kammaren Volta
University of Borås, Faculty of Textiles, Engineering and Business.
2025 (Swedish)Independent thesis Basic level (university diploma), 10 credits / 15 HE creditsStudent thesisAlternative title
Verification of EMC-chamber – Analysis of the chamber Volta (English)
Abstract [sv]

ör att säkerställa att EMC-kammaren Volta uppfyller de krav som ställs vid Emärkningscertifiering enligt ISO 11452–1 och 11452–2 för strålad immunitetstestning, har ett antal kritiska delmoment undersökts. Fokus låg på övertoner från förstärkare, absorbenternas dämpningsförmåga, strålningsbredd hos hornantenn samt en trendanalys över emissions- och immunitetsegenskaper. Projektet genomfördes i samarbete med RISE och avsåg att ge underlag för fortsatt ackreditering av testverksamhet.

Varje delmoment testades enligt gällande standarder. Övertoner från tre olika förstärkare med effekterna 30, 50 och 100 W analyserades genom mätning av fältstyrka vid grundton och upp till femte övertonen. Absorbenternas dämpning testades med både line of sight- och reflektionsmetod i frekvensområdet 0,2–18 GHz. För att bedöma hornantennens täckning mättes fältstyrkan vid flera avstånd ±75 cm från referenspunkten. Trendanalysen utfördes vidtre olika tillfällen och omfattade både strålad emission och immunitet.

Alla uppmätta övertoner låg under de krav som anges i aktuella standarder. Absorbenterna uppfyllde dämpningskraven med viss marginal, även om interferens noterades vid vissa frekvenser. Hornantennen visade sig ge tillräcklig fältstyrka vid samtliga mätpunkter, vilket innebär att endast en position behöver användas vid framtida tester. Trendanalysen visade inga avvikelser utanför godkända gränser, men en enskild frekvens vid 150 MHz bör följas upp.

Projektet visar att EMC-kammaren Volta uppfyller tekniska krav för E-märkning och att utförda mätningar ger ett starkt underlag för fortsatt ackreditering. Det är fastslaget attmätningar som normalt utförs i Volta endast behöver en position på hornantennen.Trendanalysen påvisar en godkänd och stabil kammarprestanda. Fortsatta mätningar för trendanalys rekommenderas för långsiktig verifiering.

Abstract [en]

To ensure that the EMC chamber Volta meets the requirements for E-mark certification according to ISO 11452–1 and 11452–2 for radiated immunity testing, several critical components were investigated. The focus was on harmonics from amplifiers, the attenuating performance of the absorbers, the beamwidth of the horn antenna, and a trend analysis of emission and immunity characteristics. The project was conducted in collaboration with RISE and aimed to provide supporting data for continued accreditation of testing operations.

Each component was tested in accordance with the applicable standards. Harmonics from three different amplifiers with output powers of 30, 50, and 100 W were analyzed by measuring field strength at the fundamental frequency and up to the fifth harmonic. The absorbers' attenuation was evaluated using both the line of sight- and reflection methods within the frequency range of 0.2–18 GHz. To assess the coverage of the horn antenna, field strength was measured at multiple points ±75 cm from the reference position. The trend analysis was performed at three separate occasions and included both radiated emission and immunity.

All measured harmonics were below the thresholds set by the relevant standards. The absorbers met the attenuating requirements with some margin, although interference was observed at certain frequencies. The horn antenna provided sufficient field strength at all measured points, indicating that a single test position is sufficient for future testing. The trend analysis revealed no deviations beyond accepted limits, although one frequency at 150 MHz should be monitored going forward.

The project demonstrates that the EMC chamber Volta meets the technical requirements for E-certification and that the conducted measurements provide solid evidence for continued accreditation. It is established that for the types of tests typically performed in Volta, a single horn antenna position is sufficient. The trend analysis confirms acceptable and stable chamber performance. Continued measurements are recommended to ensure long-term validation.

Place, publisher, year, edition, pages
2025.
Keywords [sv]
EMC, Trendanalys, strålningsbredd, absorbenter, övertoner, kammarvalidering.
National Category
Engineering and Technology
Identifiers
URN: urn:nbn:se:hb:diva-33705OAI: oai:DiVA.org:hb-33705DiVA, id: diva2:1972846
Subject / course
Energiteknik - Högskoleingenjör
Supervisors
Examiners
Available from: 2025-07-01 Created: 2025-06-19 Last updated: 2025-09-24Bibliographically approved

Open Access in DiVA

Arbete(7146 kB)58 downloads
File information
File name FULLTEXT01.pdfFile size 7146 kBChecksum SHA-512
3cdbf751393c307765cc3c105d0e6de6363e8adf62fa56ea80835c548d03f3cdc214e0248e950953ee2751f1b5bb9af9da6eeeff752d93473de6a5a2955a7fd1
Type fulltextMimetype application/pdf

By organisation
Faculty of Textiles, Engineering and Business
Engineering and Technology

Search outside of DiVA

GoogleGoogle Scholar
Total: 58 downloads
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

urn-nbn

Altmetric score

urn-nbn
Total: 69 hits
CiteExportLink to record
Permanent link

Direct link
Cite
Citation style
  • harvard-cite-them-right
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Other style
More styles
Language
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Other locale
More languages
Output format
  • html
  • text
  • asciidoc
  • rtf